Push saved studies to the ISCV server for remote review via Phillips Intellivue. Complete studies upload automatically — anything not labelled with patient details and operator will be deleted at upload time.

Patient numbers: An 8-digit patient number is required to view images on the server. Prefix shorter numbers with leading zeros — 1234567 becomes 01234567, 123456 becomes 00123456.

Uploading from Sonosite PX

Auto-uploads via network — just plug in


Sonosite PX uploading status
1

Plug the attached network cable into any working network socket. The Sonosite PX uploads automatically — no further steps needed.

To check upload progress: open the patient list and look at the Status column on the right.

White pause — study queued to upload

Yellow dots — currently uploading

Green tick — successfully uploaded

You can’t reorder the upload queue, but connecting often means fewer studies queued each time.

Uploading from GE Vivid IQ

Manual transfer via DICOM — 7 steps


GE Vivid IQ — network cable
1

Connect the network cable from the back of the scanner to any network port.

GE Vivid IQ — patient touchscreen
2

Tap Patient on the touchscreen (bottom-left) to open the study catalogue.

GE Vivid IQ — transfer screen
3

Tap Transfer to open the transfer screen.

Confirm From Source: Local Archive – Int HD

Confirm To Destination: DICOM Storage (HL7)

These should auto-select.

GE Vivid IQ — add to list
4

Highlight studies and tap Add to List to queue them. Repeat for additional studies.

GE Vivid IQ — copy
5

Tap Copy to start the transfer.

GE Vivid IQ — progress
6

A progress box appears at the top. A green tick next to the study means upload is complete.

Phillips Intelliview viewer