Imaging & review

Uploading to the Server

Push saved studies to the ISCV server for remote review via Phillips Intellivue. Complete studies upload automatically — anything not labelled with patient details and operator will be deleted at upload time.

Patient numbers: An 8-digit number is required to view images on the server. Prefix shorter numbers with leading zeros — 1234567 becomes 01234567, 123456 becomes 00123456.
Uploading from Sonosite PX
Auto-uploads via network — just plug in


1

Plug the attached network cable into any working network socket. The Sonosite PX uploads automatically — no further steps needed.

To check upload progress: open the patient list and look at the Status column on the right.

White pause — study queued to upload
Yellow dots — currently uploading
Green tick — successfully uploaded

You can’t reorder the upload queue, but connecting often means fewer studies queued each time.

Sonosite PX upload status screen

Uploading from GE Vivid IQ
Manual transfer via DICOM — 7 steps


1

Connect the network cable from the back of the scanner to any network port.

GE Vivid IQ — network cable

2

Tap Patient on the touchscreen (bottom-left) to open the study catalogue.

GE Vivid IQ — patient touchscreen

3

Tap Transfer to open the transfer screen. Confirm From: Local Archive – Int HD and To: DICOM Storage (HL7) — these should auto-select.

GE Vivid IQ — transfer screen

4

Highlight studies and tap Add to List to queue them. Repeat for additional studies.

GE Vivid IQ — add to list

5

Tap Copy to start the transfer.

GE Vivid IQ — copy

6

Wait for the progress indicator to complete.

GE Vivid IQ — progress

7

Images are now visible in the Phillips Intelliview viewer. Disconnect the cable when done.

Phillips Intelliview viewer